6.2 Administration

■■■■■■■■■■■■■■■■■■■■■■

1208

■■■■■■■■

tiamo 2.1 (ProcessLab)

Modifications

Comment on modification of methods

If this check box is enabled then each time that a method is modified a
modification reason and a modification comment must be entered; these
are saved in the method and shown in the method history. The reason and
comments are also recorded in the Audit Trail.

Selection

on | off

Default value

off
FDA default: on

Comment on modification of determinations

If this check box is enabled then each time that a determination is modified
a modification reason and a modification comment must be entered; these
are saved in the determination and shown in the database in the subwindow
Information on the tab Determination. The reason and comments are
also recorded in the Audit Trail.

Selection

on | off

Default value

off
FDA default: on

Comment on modification of sample data (live)

If this check box is enabled then each time that a determination is modified
a modification reason and a modification comment must be entered; these
are saved in the determination and shown in the database in the subwindow
Information on the tab Sample. The reason and comments are also recor-
ded in the Audit Trail.

Selection

on | off

Default value

off
FDA default: on

6.2.2.5

Signatures

Tab: Configuration

Tools Security settings Signatures

On the tab Signatures the parameters for the electronic signatures can be
configured.

Settings according to 21 CFR 11

[Set]

Sets the parameters to default settings for FDA-compliant work.

Inactivity delay

If this check box is enabled then the dialog window for signing will be
automatically closed when the entered time limit has expired.