2 calibrating with manual solution changing – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1011

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

995

CAL MEAS Conc
Measurement of standards for the calibration of ion-selective electrodes
(ISE electrodes).

CAL Cond
Measurement of conductivity standards for the determination of the cell
constant of conductivity measurement cells.

CAL Spec
Measuring command for wavelength calibration.

5.6.5.2

Calibrating with manual solution changing

Program part: Method

Calibration commands

The basic structure of a calibration method with manual solution exchange
looks as follows:

* Definition of the number of calibration solutions and the type of solution
exchange.

 
The request for changing the buffer or the standard must be activated in
the command CAL LOOP pH or CAL LOOP Conc, respectively. During the

Advertising