2 stepping motor - general – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1193
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
1177
Mode of operation
Stepping motors can be actuated with time controls.
Parameters
The parameters for the command STEPPING MOTOR are configured on
the following 2 tabs:
■
■
Command variables
The following command variables are generated in the method run by the
command STEPPING MOTOR and can be used in formulas under the des-
ignation 'Command name.Variable designation':
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended
5.6.11.6.2
STEPPING MOTOR - General
Tab: Method
▶ STEPPING MOTOR ▶ General
Command name
Name of the command.
Input
25 characters
Device
Device name
Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.
Selection
'Device name' | not defined
Default value
not defined