4 kft ipol - control parameters, Control parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 753

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

737

Range

0 ... 999999 s

Default value

0 s

5.6.3.5.2.4

KFT Ipol - Control parameters

Tab: Method

KFT Ipol Properties... Control parameters

Command name

Name of the command.

Input

25 characters

Endpoint

Ep with

Measured value for the endpoint.

Range

-2000.0 ... 2000.0 mV

Default value

250.0 mV

Selection

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical, fast titrations.
User
Editing the individual titration parameters which affect the titration rate.