6 Configuration

tiamo 2.1 (ProcessLab)

■■■■■■■■

1289

Selection

Record message | Display message | Stop deter-
mination

Default value

Display message

Record message
The message that the validity of the calibration has expired will be auto-
matically saved with the determination.
Display message
A message is displayed and you can choose whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity of the calibration has expired will be saved automatically with
the determination.
Stop determination
The running determination will be automatically canceled. The following
message must be confirmed with [OK].

6.7.3.4

Properties - Limits

Tab: Configuration

Sensors Edit Sensor Limits

Note

This tab is displayed only for sensors of the type pH electrode, ISE elec-
trode
or conductivity sensor.

Slope monitoring (for pH and ISE electrodes only)
on | off (Default value: off)

If this option is switched on then a check will be made to see whether the
Slope determined during calibration is within the defined limits. If it is out-
side them a corresponding message appears and you can choose whether
you will nevertheless accept the value or whether the previous value is to be
retained.

Lower limit

Lower limit value for the slope.

for pH electrode
Range

-2.147E+9 ... 2.147E+9 %

Default value

96.0 %

for ISE (+1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

55.0 mV