6 stat ph - stop conditions, Stop conditions – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 834

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

818

■■■■■■■■

tiamo 2.1 (ProcessLab)

Selection

off

Default value

off

Titrino
Range

-20.00 ... 20.00

Selection

off

Default value

off

Start rate

The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. This parameter is checked 10 s after
the start at the earliest.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

off

Default value

off

Titrino
Range

0.01 ... 150.00 mL/min

Selection

off

Default value

off

5.6.3.8.6.6

STAT pH - Stop conditions

Tab: Method

STAT pH Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off