4 digital out - digital output – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1182

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1166

■■■■■■■■

tiamo 2.1 (ProcessLab)

Functions

Moves the selected line up.

Moves the selected line down.

[New]

Opens the dialog window Digital output # in order to edit a new line (see
Chapter 5.6.11.2.4, page 1166)
.

[Properties]

Opens the dialog window Digital output # in order to edit the selected
line (see Chapter 5.6.11.2.4, page 1166).

[Delete]

Deletes the selected line.

5.6.11.2.4

DIGITAL OUT - Digital output #

Dialog: Method

DIGITAL OUT Digital outputs Properties DIGITAL OUT -

Digital output #

Port description

Selection of the digital output. The digital outputs of the selected device are
available. No port can be selected if not defined is selected in the Device
name
field.

Type

Shows the type of the digital output.

Function

Action

Selection

On | Off | Timer | Set default value

Default value

Timer

On
Sets the digital output to on after the start delay has expired.
Off
Sets the digital output to off after the start delay has expired.
Timer
Sets the digital output to on after the Start delay has expired and then
back to off after the Working time has expired.
Set default value
Sets the digital input back to the set value in the configuration.

Advertising