9 det upol - measured value window – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 549

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

533

can also be displayed in curves under the designation Extern 1…3 and used
in formulas as variables 'Command name.EX1…3'-

[New]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.10.2, page 859)
.

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 859)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.2.5.9

DET Upol - Measured value window

Dialog window: Method

DET Upol Properties... Potentiometric evalua-

tion

Evaluation with measured value window (Upol) [New]/[Proper-

ties]

Measured value window#

Measured value windows are regions on the measured value axis for which
different parameters for potentiometric evaluation can be defined. Only
those endpoints which lie within these windows and additionally meet the
parameters defined for each window will be recognized. Only one endpoint
will be recognized per window.

Lower limit

Lower limit for the measured value window.

Range

-200.0 ... 200.0 µA

Default value

-200.0 µA

Upper limit

Upper limit for the measured value window.

Range

-200.0 ... 200.0 µA

Default value

200.0 µA

EP criterion

Criterion for the recognition of equivalence points. Equivalence points
whose ERC is smaller than the set EP criterion will not be recognized.

Range

0 ... 200

Default value

5

EP recognition

Filter for the recognition of equivalence points: