2 dos - monitored dosing – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1059
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
1043
Commands
Depending on the measured value, one of the following two DOS com-
mands can be selected:
■
5.6.6.3.4DOS pH
Controlled dosing with the measured quantity pH.
■
5.6.6.3.5DOS U
Controlled dosing with the measured quantity voltage U.
5.6.6.3.2
DOS - Monitored dosing
Tab: Method
▶ Commands ▶ DOS ▶ Dosing parameters
Command name
Name of the command.
Input
25 characters
Dosing criterion
Selection
Volume/dosing rate | Volume/dosing time |
Dosing rate/dosing time
Default value
Volume/dosing rate
Specification of volume and dosing rate
You define the volume to be dosed and the desired dosing rate. The filling
times prolong the dosing process.
Specification of volume and dosing time
You define the volume and the time in which this volume is to be dosed
(particularly suitable for synthesis applications). The dosing rate is automat-
ically optimized.