2 dos - monitored dosing – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1059

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1043

Commands
Depending on the measured value, one of the following two DOS com-
mands can be selected:

5.6.6.3.4DOS pH
Controlled dosing with the measured quantity pH.

5.6.6.3.5DOS U
Controlled dosing with the measured quantity voltage U.

5.6.6.3.2

DOS - Monitored dosing

Tab: Method

Commands DOS Dosing parameters

Command name

Name of the command.

Input

25 characters

Dosing criterion

Selection

Volume/dosing rate | Volume/dosing time |
Dosing rate/dosing time

Default value

Volume/dosing rate

Specification of volume and dosing rate
You define the volume to be dosed and the desired dosing rate. The filling
times prolong the dosing process.

Specification of volume and dosing time
You define the volume and the time in which this volume is to be dosed
(particularly suitable for synthesis applications). The dosing rate is automat-
ically optimized.