2 stdadd auto - general/hardware – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1003

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

987

Identifi-
cation

Description

.ENP

Electrode zero point calculated from the calibration

.ETE

End temperature (temperature following processing of the
command) in °C

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

.IME

Initial measured value (measured value before start condi-
tions are processed) in the unit of the measured value

.ITE

Initial temperature (temperature before start conditions are
processed) in °C

.MTE

Temperature measurement with sensor; 1 = on, 0 = off)

.RES

Calculated result of the standard addition in the selected
unit

.SLO

Electrode slope calculated from the calibration

.STY

Type of stop with which the command was stopped: 1 =
normal
;0 = manual or after error

.TITER

Titer value of the solution used for the command

.VAR

Variance of the calculated result of the standard addition in
the selected unit

.VOL

Dosed volume

5.6.4.16.2

STDADD auto - General/Hardware

Tab: Method

STDADD auto Properties... General/Hardware

Command name

Name of the command.

Input

25 characters

The general parameters for the control device, the dosing device, the sensor
and the stirrer are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.

Selection

'Device name' | not defined

Default value

not defined