8.4 Manual Control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1478

■■■■■■■■

tiamo 2.1 (ProcessLab)

Rack code

Shows the number of positions on the rack attached. If no rack is attached,
-------- is displayed.

Number of positions

Shows the number of positions on the rack attached. If no rack is attached,
-------- is displayed.

Shift rate

Shift rate for manual control of the sample changer.

Range

5 ... 20 °/s

Default value

20 °/s

Lift rate

Lift rate for manual control on the selected tower.

Range

5 ... 25 mm/s

Default value

25 mm/s

Swing rate

Swing rate for manual control of the Swing Head on the sample changer.

Range

10 ... 55 °/s

Default value

20 °/s

[Initialize rack]

The rack in position is initialized.

Note

When initializing the rack, the following actions are carried out:

– Move rack in position to read out rack code.

– Transfer rack data for the rack code to sample changer.

– Move lifts upwards to 0 mm

– Move robotic arm back.

8.4.4.3

Sample changer - Move

Program part: Manual Control

Here the moving to desired rack, lift or robotic arm position can be triggered
manually.

Rack position
Set rack position.