Setting the Recorder

Scan Interval/Memory

Set these parameters when the data type is set to "EVENT & DISP" or "EVENT"

Check the channels you wish to sample

Scan interval

The selectable scan intervals vary depending on the model as follows:

DX102, DX104, DX204, DX208, DX204C, and DX208C :

125 ms and 250 ms

DX106, DX112, DX210, DX220, and DX230 :

1 s and 2 s

MV102, MV104, MV204, and MV208 :

125 ms and 250 ms

MV106, MV112, MV220, and MV230 :

1 s and 2 s

A/D Integrate

100 ms can be selected only when the scan interval is set to 2 s.

Memory Sample (save method of measured/computed data)

• Number of blocks

When the data type is [EVENT], select 1, 2, 4, 8, or 16.

When the data type is [EVENT&DISP], select 1, 2, or 4.

• Pre-Trigger Length
If 0% is selected, the event file will entirely consist of data after the trigger. If 100% is

selected, the event file will entirely consist of data before the trigger.

• Memory Sample

Select the channels that are to be saved to the memory.

Note

If [Save] is set to [Manual], the data directory is created at a location that cannot be managed

by the DAQ Desktop. Therefore, the DAQ Desktop cannot be used to handle data files in that

directory.

7.6 Setting DX100, DX200, MV100, MV200