Yokogawa DAQWORX User Manual

Page 298

Advertising
background image

7-122

IM WX101-01E

7.8 Setting MV1000/MV2000

Memory Sampling

Turn [ON] (sample) or [OFF] (do not sample).

Display Zone (Zone L and U)

You can select the range of the screen in which the waveform of each channel is to be

displayed.

Specify positions (%) on the display scale for the upper and lower limits.

The conditions for setting the zones are as follows:

• Range: 0% to 100%

The lower limit L must be less than the upper limit

• The difference between the lower and upper limits is at least 5%.

Graph

For details, see section 5.7 in the MV1000/MV2000 User’s Manual (IM MV1000-01E).

Scale display position

Select the scale display position on the trend display from 1 to 10 for the MV2000 or from

1 to 6 for the MV1000. Select [OFF] if you do not wish to display the scale.

Scale divide position

Select the number of main scale marks on the trend display from 4 to 12 and C10.

C10:

The scale is equally divided into 10 sections by main scale marks, and scale

values are indicated at 0, 30, 50, 70, and 100% positions on the trend display.

Bar display position

Select [Normal], [Center], [Lower], or [Upper].

Bar divide number

Select number of divisions of the scale on the bar graph display.

Partial (Partial Expanded Display)

Bound position (%)

Set the boundary for the partial expanded display. The range is from 1 to 99%.

Boundary

Set the value that is to be the boundary between the reduced section and the expanded

section in the range of “minimum span value + 1 digit to maximum span value – 1 digit.”

For channels that are set to scaling, the selectable range is “minimum scale value + 1

digit to maximum scale value – 1 digit.”

Example: Input range: –6 V to 6V. Bound position: 30. Boundary: 0

The –6 V to 0 V range is displayed in the 0% to 30% range, and the 0 V to 6

V range is displayed in the 30% to 100% range.

The conditions used to set the boundary vary depending on the measurement and

computation channels as follows:

• Measurement channel

When SCALE and SQRT are not used: Span L < boundary < span U

When SCALE and SQRT are used:

Scale L < boundary < scale U

• Computation channel

Span L < boundary < span U

Note

For the MV1000/MV2000, this is when [Partial] is turned [ON] under [Detail Setting] in the [Basic

Setting] tab.