Index – Yokogawa DAQWORX User Manual
Page 59

2-39
IM WX101-01E
1
2
3
4
5
6
7
8
9
Index
Configuring the DAQLOGGER
4. If necessary, set whether or not to measure in zero start adjustment mode.
In zero start adjustment mode, when the scan interval is less than a minute (1 to 59),
the start time of the scanning operation is adjusted so that the operation takes place
at 00 sec every minute. If the scan interval is greater than or equal to one minute (60
to 3600), the scanning operation is adjusted so that the operation takes place at 00
min 00 sec every hour.
The zero start adjustment mode is valid for scan intervals which are a divisor of 60
when it is less than a minute, and for scan intervals which are a divisor of 3600 when
it is greater than or equal to one minute.
5. If necessary, set whether or not to record in the zero start adjustment mode.
The zero start adjustment for recording is valid when the zero start adjustment for
scanning is enabled.
When the record interval (scan interval x recording rate) is less than a minute (1 to
59), and is a divisor of 60, the start time of recording is adjusted so that recording
starts at 00 s every minute. If the record interval is greater than or equal to one
minute and less than one hour (60 to 3599), and is a divisor of 3600, the start time
of recording is adjusted so that recording starts at 00 min 00 s every hour. If the
record interval is greater than or equal to one hour and less than or equal to one day
(3600 to 86400), and is a divisor of 86400, the start time of recording is adjusted so
that recording starts at 00 hour 00 min 00 s everyday.
6. In the “Start” box, select the start condition, “On Record” or “Fixed time.”
If you select “On Record,” the recording operation starts when the “RECORD” button
is clicked. If you select “Fixed time,” the recording operation starts automatically at
the specified time.
7. In the “End” box, select the end condition, None, Fixed time, or Data count.
If you select “None,” the recording operation stops when the “STOP” button is
clicked. If you select “Fixed time,” the recording operation stops automatically at the
specified time. If you select “Data count,” the operation stops when the specified
number of data points have been recorded.
8. Click the “Select folder” button to specify the folder used to save the recording
results (data file).
The “Browse for Folder” dialog box appears.
9. Specify the folder to save the data file, then click “OK.”
Note
Report files are also saved to this folder.
2.9 Configuring the Logger