Setting the Recorder

Ref. CH

The reference channel for difference computation.

Square Root

Computes and displays the square root of the input. This setting can be used only when

the input mode is set to VOLT (voltage). As necessary, set the span, scale, and unit.

Unit

Enter the unit using up to six characters.

Low-cut (Can be set when the mode is 1-5V, and when the mode is VOLT with
square root (SQRT) selected. )

Select [ON] to use the low-cut function.

Low-cut value (Can be set when the mode is VOLT with square root (SQRT)
selected.)

Set the low-cut value in the range of 0.0% to 5.0% of the input span.

Input value

Low-cut value

Result of square root computation

Measured value