Yokogawa DAQWORX User Manual

Page 342

Advertising
background image

7-166

IM WX101-01E

Mode (input type)

Scale ON/OFF

Power monitor wiring

Power monitor input range

Scale left

Scale right

Span left

Span right

Measurement range

Scale unit

Reference channel for Delta or RRJC mode

Selecting the Range and Type of Input

The following types of input can be selected. The default setting is VOLT.

• SKIP

Measurement, data saving and display (except for page display) will not be carried

out. Measurement, data saving and display will be carried out for the next channel

whose input type is not set to SKIP.

• VOLT (DC voltage)

This input type can be selected from 20mV, 60mV, 200mV, 2V, 6V, 20V and 50V.

Refer to chapter 14 for the measurement range of each setting. The default setting is

2V.

• TC (thermocouple)

This input type can be selected from R, S, B, K, E, J, T, N, W, L, U and KP (KPvsAu7Fe).

Refer to chapter 14 for the measurement range of each setting. The default setting is R.

• RTD (resistance temperature detector)

This input type can be selected from PT1 (Pt100 1mA), PT2 (Pt100 2mA), JPT1 (JPt100

1mA), JPT2 (JPt100 2mA), PT50 (Pt50 2mA), NI1 (Ni100 1mA SAMA), NI2 (Ni100

1mA DIN), NI3 (Ni120 1mA), CU1 (Cu10 GE), CU2 (Cu10 L&N), CU3 (Cu10 WEED),

CU4 (Cu10 BAILEY), PT1S (Pt100 1mA high resolution), PT2S (Pt100 2mA high

resolution), JPT1S (JPt100 1mA high resolution), JPT2S (JPt100 2mA high resolution)

and J263B (J263*B). Refer to chapter 14 for the measurement range of each setting.

The default setting is PT1.

• DI (select LEVL (voltage level) or CONT (contact)

For LEVL, a voltage of less than approx. 2.4 V will be recognized as “0 (OFF),”

whereas a voltage of approx. 2.4 V or more (max. allowable voltage is up to ±60 VDC)

will be recognized as “1 (ON).”

For CONT, an open, externally connected contact to which no voltage is applied, will

be recognized as “0 (OFF),” whereas a closed contact will be recognized as “1 (ON).”

The default setting is LEVL.

• mA (DC current, mA input module)

This input can be set only for mA input channels. The measuring range is -20 mA to

20 mA.

7.9 Setting DARWIN

Advertising
This manual is related to the following products: