Index – Yokogawa DAQWORX User Manual

Page 441

Advertising
background image

7-265

IM WX101-01E

1

2

3

4

5

6

7

8

9

Index

Setting the Recorder

3. When the setting parameters have been sent, a “Communication Message” dialog

box appears.

Read the message and click “OK.”

Note

• Communication is not possible of the scanning or the recording is in progress.

• With the VR series, the settings data of “SETUP” cannot be sent to the recorder.

• With the DX/MV/CX series, the settings of network in the “Setup” tab panel cannot be sent

to the recorder, except for the setting of memory data output destination.

• Use the “File Export” function to set both series.

• Before connecting the hardware, check if Module Configuration matches the actual module

construction.

• If a file remains in the internal RAM disk of the DC100, it is not possible to send settings

to its internal RAM disk. Initialize the internal DC100 RAM disk before sending data to the

DC100, or initialize the DC100 using this software. In that case, make sure to first copy

important files.

• If files remain on the DC100 internal RAM disk, when sending data to the DC100 the writing

action does not change automatically to SINGLE or REPEAT, when it was earlier set to

NONE on the DC100, and likewise, not to NONE if it is set to SINGLE or REPEAT on the

DC100.

• If files remain on the DC100 internal RAM disk, when sending data to the DC100 , the

periodic file type does not change automatically to Meas or Report, when it was earlier set to

NONE on the DC100, and likewise, not to NONE if it is set to Meas or Report on the DC100.

• If data are sent, during computation, reporting, or (for DC only) data writing, those actions

will be canceled.

• If data are sent, during balancing for strain input or (for DR only) data writing to the internal

RAM disk, a communication error will be detected.

• If only altered settings are send event/action, Math, Report, or Memory Pause, LEVEL

actions, will not be canceled.

• When using Ethernet communication, error detection may take a few minutes.

7.11 Writing Setting Parameters to the Recorder

Advertising
This manual is related to the following products: