7-184

IM WX101-01E

Note

If data are written using TRIG, the saving operation is also required to start using the START

key or through the event/action function.

Selection of Writing Data (WRITE DATA)

Select the data to be written from the following:

• All :

All measured/computed data are written.

• Alarm : Measured/computed data in all channels are written only when an alarm is

generated.

Setting of File Name (WRITE NAME)

Select the specifying method of a file name from the following:
• Set :

Freely set by the user (up to 5 characters). Characters that can be used for
file names are only those that can be selected from the display; no blanks
allowed. All letters will automatically be capitalized.

• Date : Automatically set based on the date.

Writing Period

• The period in which data are written can be selected from the following:

• INTVL : The same as the measuring period. Measured/computed data are saved

without thinning-out.

• 1, 2, 3, 4, 5 , 6, 10, 12, 15 , 20 , 30 seconds, or, 1, 2, 3, 4, 5, 10, 30, or 60 minutes :

Writing is done with the set period.

However, only those values which are multiples of the measuring period can be

selected.

• LOGIC : Only effective if event/action is set to action DATA_WR. Every time an event

is generated, the data item for 1 scan (1 for each channel) is written.

• When the measuring period is changed, the writing period will not equal the measuring

period or its integral multiples, and is subsequently adjusted to equate with the

measuring period.

Built-in RAM Disk Usage (WRITE MEMORY)

• Select the built-in RAM disk usage from the following:

• LENGTH : Determines file capacity by specifying the number of data items per ch.
• DIV2, DIV4, DIV8, or DIV16 : Use the RAM disk by dividing the disk area into 2, 4, 8

or 16 parts. If the memory capacity is 4 MB, 2 cannot

be used.

• If a file is saved in the built-in RAM disk, the usage of the built-in RAM disk cannot be

changed. For this purpose, initialize the RAM disk or change the setting after deleting
all the files. In that case, necessary files must be saved on a floppy disk.

Writing Data Length (WRITE LENGTH)

• If WRITE MEMORY is set to LENGTH, set the number of data items to be written per

channel, selecting from the following: 10, 20, 30, 40, 50, 100, 200, 300, 400, 500, 1k,

2k, 3k, 4k, 5k, 10k, 20k, 30k, 40k, 50k, or 100k

• However, a data length exceeding a file size of 1 MB cannot be selected depending

on the number of channels.

Note

When using the computation channel, do not write measured/computed data in an application

where the measurement interval is shorter than 0.5 sec, the writing interval is set to INTVL, or

the data length is shorter than 50, if:

• the writing action is in the REPEAT mode.

• writing starts using the event/action mode.

7.9 Setting DARWIN