Calibration correction – Yokogawa DAQWORX User Manual

Page 296

Advertising
background image

7-120

IM WX101-01E

7.8 Setting MV1000/MV2000

Low-cut (Can be set when the mode is 1-5V, and when the mode is VOLT with
square root (SQRT) selected. )

Select [ON] to use the low-cut function.

Low-cut value (Can be set when the mode is VOLT with square root (SQRT)
selected.)

Set the low-cut value in the range of 0.0% to 5.0% of the input span.

Input value

Low-cut value

Result of square root computation

Measured value

Calibration Correction

Set the input and output values for the calibration correction. The number of set points

(including the start and end points) can be specified in the range 2 to 16.

Click to delete the selected row.

Click to add set points (rows) to the number of calibration set points.

Input

Number of set points

Calibration set points

Y5
Y4

Y3
Y2

Y1

Output

X1 X2 X3 X4

X5

Condition

X1 < X2 < X3 < X4 < X5

Selectable Range of Input and Output Values
• Channels on which linear scaling is specified

–30000 to 30000 (the decimal place is the same setting as the scale value)

• Other channels

Value in the measurable range of the selected range

Example: –2.0000 to 2.0000 for 2 V range

Advertising
This manual is related to the following products: