background image

7-239

IM WX101-01E

1

2

3

4

5

6

7

8

9

Index

Setting the Recorder

Memory Sample (save method of measured/computed data)

• Save

Select the save method of internal memory data to an external storage media from

[Auto] or [Manual].

Manual: Inserting the external storage media into the drive and closing the cover

displays the “Save confirmation” message, allowing data to be saved.
When the operation is complete, remove the external storage media from

the drive, so that the next set of data save operation can be performed. You

can select whether to save all of the data from internal memory or only to

update the data still not saved to an external storage media.

Auto:

If an external storage media is always in the drive, data is saved

automatically at a preset interval.

• Data

Select the data to be written to internal memory from the following: [DISPLAY]

(displayed data only), [EVENT & DISP] (event data and displayed data), or [EVENT]

(event data only).

• Event Data Sampling Rate

Select the interval at which event data is saved from the following: [1s], [2s], [5s], [10s],

[30s], [60s], [120s], [300s], or [600s].

• Event Data Sampling Mode

Select [Free], [Trigger] or [Rotate].

• Block

When the data type is [EVENT], select 1, 2, 4, 8, or 16.

When the data type is [EVENT & DISP], select 1, 2, or 4.

• Data Length

Set the interval corresponding to the amount of data (data length) that can be written

as a block of the event data storage region. The data length that can be set depends

on the event data sampling rate. It also depends on the block setting and number of

Meas and Math channels.

• Pre-Trigger Length
If 0 is selected, the event file entirely consists of data after the trigger. If 100 is

selected, the event file entirely consists of data before the trigger.

• Manual Trigger

When applying triggers with keys, select [ON].

• External Trigger

When applying trigger signals by remote input, select [ON].

• Alarm Trigger

When applying alarms as triggers, select [ON].

• Sampling