Setting constants, Selecting events – Yokogawa DAQWORX User Manual

Page 352

Advertising
background image

7-176

IM WX101-01E

Setting Constants

Constants are used for computation equation on models with computation function (/M).

Up to 60 constants (K01 to K60) can be set (Thirty constants (K01 to K30) for the

standalone type).

The number of significant digits is 5 excluding the decimal point. If an exponent is used,

the mantissa and exponent must consist of 5 digits and 2 digits, respectively.
• Configurable ranges:

–1.0000E+35 to –1.0000E-35

0

1.0000E-35 to 1.0000E+35

Constants tab

Selecting Events

Event/Action and Match Time settings are possible for the DA100, if the (/M1) Math

option is being used, or if a pulse module is attached.

For DA100 and DC100, Group settings are possible if the Math option is being used.

Group settings

Match time settings

Action settings

Edge/Level settings

Event settings

Remote, Relay, Timer, and Match Time number settings

Event

Select any event from among the following:

NONE — releases the event/action setting.

REMOTE

• This menu is displayed only with the DI/DO modules that are recognized as system

modules.

• When a remote control signal is applied, the action is executed.

• There are 12 control signals available. Select any control signal from these numbers

for event setting. Remote control signal terminal numbers are applicable for this

selection.

7.9 Setting DARWIN