3

Operation

PC Control / Touch Control

97

The dialog in which you can enter the templates for sample

identifications is opened automatically with [New]. For an existing

template this dialog can be opened with [Edit]. Existing sample

identifications can be removed from the list with [Delete].

Exit the dialog with [Back] and activate the Use sample

identification list checkbox so that you can select from the

templates when entering the sample data.

3.14.2 Sample assignment table

If you use the sample assignment table then you can be sure that the

samples will be processed with the correct method. In the sample

assignment table a sample identification is assigned to a particular

method. Samples with an identification which appears in this table can

only be processed by the assigned method. This means that no

mistakes can occur. The user enters the sample identification and at

the START of the determination the correct method is loaded

automatically.

Open the dialog Templates/Sample data.

Use [Sample assignment table] to open the list of sample

identifications with their assigned methods.

You can define new sample assignments with [New].