Activate the checkbox Spooler in order that the Touch Control is

not blocked by printing out a report so that you can continue

working.

Note!

The Touch Control prints the reports with a fixed resolution of 300 dpi.

If you are using a printer with a resolution of 360 dpi (or a multiple,

e. g. Epson) then the text will be printed out somewhat smaller than

with printers with a resolution of 300 dpi (or a multiple, e. g. Canon or

HP).

3.10.8 Balance

Connecting a balance to the Titrando or PC is described in the

Installation Instructions for the Titrando. In the device manager you

must configure the serial interface used to connect the balance.

Select the Balance type.

The following table shows the balance type to be selected for various

balance models:

Balance

Balance type

AND

AND

Mettler AB, AG, AM, PM, XP, XS

Mettler

Mettler AT

Mettler AT

Mettler AX, MX, UMX, PG, AB-S

Mettler AX

Ohaus

Voyager, Explorer, Analytical Plus

Ohaus

Precisa

Precisa

Sartorius

Sartorius

Shimadzu BX, BW

Shimadzu

PC Control:

Under RS-232 select the RS-232 interface (COM) of the PC to which

the balance is connected. All the COM interfaces of the PC will be

shown.

Add, if not present already, the virtual device Serial ports to the

device manager and open the list with the available RS-232

interfaces with [Edit]. Select the interface to which the balance is

connected. Open the dialog with the port parameters with [Edit]

and configure these. The settings on the balance must agree with

these settings here. A description of the parameters can be found in

the online help.