Titration sequence for set – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 223

Advertising
background image

4

Parameters

PC Control / Touch Control

211

Titration sequence for SET

1st start of titration

command

Checks whether work is to be carried out with or

without conditioning.

(Switch on stirrer)

The stirrer is switched on after the start.

Preconditioning

(if activated)

The titration cell solution is titrated to the end-

point and kept there continuously.

2nd start of titration

command

The method sequence itself starts.

Measure initial meas-

ured value

The initial measured value is measured.

(Start conditions:

Pause 1

Start volume

Pause 2)

Pause interval 1 is allowed to elapse. The start

volume is added. No measuring points are en-

tered in the list of measuring points. Pause inter-

val 2 is allowed to elapse.

(Extraction time)

Titration with test up

to stop when stop

criterion or the stop

condition is reached

Titration is carried out to the first and then to the

second endpoint.

If the extraction time has not elapsed when the

(first) endpoint has been reached then it will be

allowed to elapse and the titration (to the first

endpoint) will only be concluded when the ex-

traction time has elapsed.

(Switch off stirrer)

The stirrer is switched off if there is no recondi-

tioning.

Fill dosing

cylinder

After the end of the titration the dosing cylinder is

refilled automatically.

Reconditioning

The titration cell solution is kept continuously at

the endpoint.

Optional steps are shown in brackets.