3.9 Sensors

52

PC Control / Touch Control

3.9.3 Editing sensor data

Open the dialog Sensors/Edit by either configuring a new sensor

(see Section 3.9.1) or selecting an existing sensor from the list and

modifying it with[Edit].

Enter an unambiguous name for the sensor under Sensor. For ISE

electrodes the ion whose concentration can be determined with the

sensor is also shown.

The Order number and Serial number are read out automatically

in case of intelligent sensors. The data can be entered manually for

non-intelligent sensors.

You can enter a Comment for each sensor.

3.9.4 Calibration data (for pH and ISE electrodes only)

Detailed information about calibrations is given under Edit

sensor/Calibration data. For pH electrodes the Slope is given in %

and the electrode zero point pH(0) is shown. For ISE electrodes the

Slope is given in mV; the electrode zero point E(0), the blank

concentration c(blank) and the variance are also shown. The variance

is only calculated, if the calibration has been carried out with at least

four standards. The calibration date is entered automatically after a

calibration has been carried out (see Section 4.7).
Under Calibration temperature the temperature at which the

calibration was carried out is shown. If the temperature was measured

and entered manually during the calibration then this is indicated by

manual. If the temperature was measured with a connected

temperature sensor then the type of sensor (Pt1000 or NTC) will be

shown. The Calibration date is entered automatically together with

the time. When a sensor is entered into the list for the first time the date

of entry will be given as the calibration date.
Under Calibration method the method used to calibrate the sensor is

shown. If the calibration data has been entered manually then manual

will be shown here. The user who carried out the calibration or entered

the calibration data manually is shown under User.