Function

: Selects the possible working steps.

With Dosing the given volume is added. Automatic filling takes

place neither before nor after the command.
Fill

the cylinder can take place from a freely selectable port. The

valve disk then remains at the selected port.
With Aspirate liquid is aspirated via the given port. As with Dosing,

the cylinder is not automatically filled either before or after the

command. It should be possible for the aspirated volume to be re-

moved with a single piston stroke.
With the Eject function the whole cylinder content is ejected via the

given port. In contrast to the function "End volume" the piston is

pressed down to the stop, i.e. past the maximum volume mark. This

command should only be used for preparing the dosing unit before

the pipetting sequence.
With the command Exchange position the cylinder is filled via the

given port. This means that, for example, air can be aspirated via

Port 4. The valve disk is then rotated to Port 2 and the dosing drive

can be removed from the dosing unit.
With Change port the valve disk is only moved to the given port; no

piston movement takes place.
As the dosing units are exchangeable, the coupling of the Dosino

piston rod (spindle) has a slight mechanical tolerance that can be

noticed when the piston changes its direction of movement. This

tolerance can be compensated with the command Compensate. A

short piston movement is first made in the same direction as the

previous movement; this is followed by a piston movement in the

reverse direction.
The whole cylinder content is ejected via the selected port with End

volume

. The piston is lowered to the maximum volume mark i.e. un-

til 10'000 pulses have been dosed. This command should be used

for pipetting functions to empty the cylinder.

In-/Outlet: Selects the port, with which the liquid handling com-

mand should be carried out.

Volume: Here you enter the volume to be processed.
Rate: Rate at which the commands Dosing, Fill, Aspirate, Eject,

Exchange position

and End volume are carried out. Aspirating

and ejecting the sample should be done with a rate < 10 mL/min.

The item Control device is mentioned in Section 4.10.1, the parame-

ters for the Dosing device are described in Section 4.1.5.
A schematic command sequence is explained in Section 6.5.3.