3.14 Templates

100

PC Control / Touch Control

3.14.4 Input lines

Binary patterns can be defined for scanning remote lines with the

SCAN command (see Section 4.11.1) during the determination

sequence; these can then be selected when editing the parameters.

Open the dialog Templates/Input lines.

You can define new templates with [New]. Existing templates can be

removed from the list with [Delete] or edited with [Edit].

List of predefined input lines
Parameter Signal

pattern

Function

Cond OK

******1*

waits for "Cond OK" condition of device 1

End1

****1***

waits for the EOD impulse from device 1 (Titrino,

Titrando)

End2

*1******

waits for the EOD impulse from device 2

EndMeter

***11***

waits for the EOD impulse from ion meter or pH-Meter

(stirrer 1 will be switched on during the waiting period)

Ready*

**1****1

waits for "ready" condition of devices 1 and 2 (Titrino,

Titrando)

Ready1

*******1

waits for "ready" condition of device 1

Ready2

**1*****

waits for "ready" condition of device 2

Sample ready

***1****

waits for an impulse to continue e. g. of a connected

Sample Processor as soon as it is ready

With the parameter Ready* it is possible to scan the condition of

devices working in parallel. In this case the 'Ready' line of both devices

must be statically set (i.e. constantly) at the end of a determination.

Devices which only transmit a brief impulse at the end of e. g. a

measurement cannot be controlled in parallel.

Use [Edit] to open the edit dialog for the template Cond OK.