Fill, Prepare – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 194

Advertising
background image

3.24 Manual control

182

PC Control / Touch Control

The specified volume is shown above the volume that is already dosed.

If the Fill automatically checkbox is deactivated then the button

[Fill] is available additionally.
[Stop] is used to stop the addition. You can use [Hold] to interrupt the

addition and continue it with [Continue].

Fill
You can fill the dosing cylinder manually with the [Fill] function. The

piston and valve positions are shown during the filling process. Filling is

carried out at the maximum filling rate.

Prepare
The [Prepare] function is used to rinse the cylinder and tubing of the

exchange unit or dosing unit and to fill bubble-free. You should carry

out this function before the first determination or once per day.
The parameters defined for the titrant under System/Titrants for

preparing the exchange or dosing unit and the tubing parameters will

be used for rinsing (see Section 3.8.6). For every dosing device you can

define in the Device manager when a message is to be shown to

remind you that the dosing device has to be prepared (see

Section 3.10).

Start the preparation process with [Prepare]. The display switches

automatically to the live display.