3.10 Device manager

70

PC Control / Touch Control

Touch Control:

Under RS-232 select the RS-232 interface at the RS-232/USB Box

which is used to connect the balance.

Select the RS-232/USB Box in the device manager and open the list

with the available RS-232 interfaces with [Edit]. Select the

interface to which the balance is connected. Open the dialog with

the port parameters with [Edit] and configure these. The settings

on the balance must agree with these settings here. A description of

the parameters can be found in the online help.

Note!

In case of communication problems you can try using the software

handshake.

The sample size is transferred as a number with up to ten places, sign

and decimal point. Units transmitted by the balance will also be

transferred. Control characters will not be transferred. With some

balances it is also possible to transmit the sample identification and

method from the balance as well as the sample size. When editing the

sample data in the sample data silo all the data can be transmitted

from the balance (see Section 3.19.3). In some cases a special input

unit may be necessary; this can be supplied by the balance

manufacturer.
In order to do this the following addresses for the identifications and

methods must first be selected:

Balance

Method

Identification 1 Identification 2

Sartorius

(input unit)

METH or 27

ID.1 or 26

ID.2 or 24

Mettler AT

(input unit)

D (Mthd)

C (ID#1)

B (ID#2)

Mettler AX

Label for ID

which

contains the

method name:

METHOD

Label for ID

which

contains

Identification 1

:

ID1

Label for ID

which

contains

Identification 2

:

ID2