3.7 System settings

32

PC Control / Touch Control

If you activate Methods then when a modified method is being saved a

reason for the modification will be demanded. Results are altered when

a determination is recalculated. If you require a reason to be given in

such cases then activate the Determinations checkbox. The reasons

are only demanded when login and password is used.
These reasons are documented in the audit trail (see Section 3.7.8)

together with the alteration. The reason for the last modification is

shown in the properties of the method or determination.

You should now edit the password options. You can define a

minimum length for the passwords to be used. This will be

checked when a user enters a new password. After the maximum

Number of entry attempts during login the user will be

automatically deactivated and can only be reactivated by a user

with administrator rights. The use of certain Special characters in

the password can be demanded (see online help). You can also

define a time limit after which the password must be altered.

Changing a password, it is not possible to choose one which was

already used before.

In the Titrando Compliance Guide there is a step by step description

of the necessary procedure for complying with FDA Guideline 21 CFR

Part 11.