5.1 Troubleshooting

292

PC Control / Touch Control

Problem

Possible cause

Remedy

Cylinder volume of

exchange/dosing unit is too

large.

Use an exchange/dosing unit with a

smaller cylinder volume (smaller

volume increment per pulse).

Electrode responds to

slowly.

Replace the electrode.

Time for reaching

the control point is

too long.

Unfavorable control

parameters.

- Select a smaller Dynamics under

User-defined parameters.

- Increase the Min. rate under

User-defined parameters.

- Increase the Max. rate under

User-defined parameters.

5.1.7 Sensor

data

Problem

Possible cause

Remedy

After calibration the

calibration data is

not stored together

with the sensor data.

In the method the correct

Sensor has not been

selected in the CAL

command.

Take care that the sensor to be

calibrated has been selected under

Edit command/Sensor (see

Section 4.1.5).

5.1.8 Print

Problem

Possible cause

Remedy

The [Print] fixed key is

deactivated.

Activate the [Print] fixed key

under Dialog options/

Fixed keys (see Section 3.7.3).

The [Print] fixed key

is inactive.

The system is in the BUSY

status, i.e. a determination is

running.

Wait until the determination is

finished.

PC Control: empty

pages are produced

when printing out

the result silo report

or the statistics

overview.

Landscape format has

been selected under

orientation in the Windows-

specific printer settings.

in the Windows-specific printer

settings select portrait format as

the printout is automatically made

in landscape format.