6.5.3 Pipetting

sequences

Principle
Measuring out the liquid to be pipetted takes place in the pipetting tub-

ing, which is filled with a carrier liquid (preferably water). In order to

avoid carryover or mixing an air bubble (separating bubble) must be in-

serted between the sample and the carrier liquid. The tubing tip should

be drawn out by hand to form a fine point.
A Dosino drive is used to transfer the liquids to the tubing. The pipetting

tubing is attached to Port 1 of the dosing unit. The dosing unit is

mounted on a storage bottle containing the carrier liquid. This can be

refilled via filling Port 2 of the dosing unit. The carrier liquid can also be

used for rinsing and diluting the sample.
Numerous pipetting sequences are possible. The following conditions

must always be observed:
• The sample must always be measured out in the pipetting tubing; it

must never enter the dosing unit cylinder.

• The sample should always be aspirated after the dosing cylinder

has been ejected.

• The difference in level between the surface of the sample liquid and

that of the carrier liquid in the dosing cylinder should be as small as

possible.

• Aqueous sample solutions can be pipetted without having to rinse

the tubing. Non-aqueous samples require the pipetting tubing to be

rinsed with carrier liquid (same solvent as the sample) to avoid

carryover.

Only the most necessary commands are mentioned in the following

command sequence. A complete command sequence is defined in the

method template "SP Pipetting Ext. Position" (see Section 3.15.2).

Preparing the dosing unit
The dosing unit must first be prepared, i.e. the filling tubing and pipet-

ting tubing must be rinsed and completely filled. The preparation step is

best carried out at an external position. The cylinder contents must then

be ejected. This command sequence is best combined to a subse-

quence. This preparation step is only necessary at the beginning of a

sample series. Therefore the subsequence is defined as Start se-

quence (see Section 4.9.6).
Command sequence:

MOVE

move to external position

LIFT

move to work position

PREP

prepare dosing device

LQH

eject

cylinder

These steps should be carried out before every sample series involving

pipetting. Before each pipetting process the dosing unit cylinder con-

tents must be completely ejected. This is the only way to achieve re-

producible, accurate pipetting.