3

Operation

PC Control / Touch Control

61

head cannot move to any lower position than the value entered here

(0 mm = upper lift stop position).

• Under Min. beaker radius you can enter the smallest permissible

sample beaker radius for automatic and manual operation. This is

also a safety setting in order to prevent a wide, fully equipped

titration head from trying to enter a narrow sample vessel. Before

the lift is moved to the work position, a comparison will be made

between the value entered here and the beaker radius given in the

rack table (see p. 63) and, if necessary, an error message will be

displayed. There will be no check if no radius has been defined.

• The Lift rate can be set for manual operation.
• The rotation increment is the angle about which the sample rack

should be rotated relatively to the momentary position. This rotation

increment is used for the settings + rotate and – rotate in the

MOVE command.

• The axial distance is the distance between the axis of rotation of

the sample rack and the swing axis of the robotic arm. This value

depends on the model of the USB Sample Processor and is

specified in the online help.

Swing Head

The assigned swing angle and the specific work position are shown for

each external position in the properties of the Swing Head.
The details of the dialogs [Ext. pos. 1-4], [Robotic arm] and

[Adjustm. data] are described on the following pages.
[Edit] can be used to assign a specific work position and a swing

angle to each external position.