Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 232

Advertising
background image

4.1 Titrations

220

PC Control / Touch Control

Upper limit

Lower limit

Stop

Measured

quantity

Time

Fig. 4.11: Action "Exit method" or "Skip command"

When selecting Hold, the addition of the reagent is interrupted until

the method sequence is continued manually with [Continue]. If the

monitored quantity is still outside the limits (including hysteresis), it

is not possible to continue the method sequence.

Upper limit

Lower limit

Hold

Measured

quantity

Time

[Continue]

Fig. 4.12: Action "Hold"

When selecting Wait, the addition of the reagent is interrupted until

the monitored quantity is within the limits (including hysteresis)

again. The dosing is continued automatically.