background image

5 Troubleshooting – Maintenance

PC Control / Touch Control

289

5.1.4 SET

titrations

Problem

Possible cause

Remedy

Max. rate too low.

Increase the min. rate under

Control parameters/User-def.

parameters.

Dosing at the end

takes too long and

the increments are

too small.

Unfavorable Stop

criterion.

Try increasing the Stop drift or

select a shorter Delay time.

The titration is not controlled

properly, i.e. individual

increments are not added at

the end.

- Select the Titration rate slow

under Control parameters.

- Select a larger Dynamics under

Control parameters/User-def.

parameters.

- Reduce the Max. rate under

Control parameters/User-def.

parameters.

- Reduce the Min. rate under

Control parameters/User-def.

parameters.

- Ensure that the solution is better

mixed and that the arrangement of

electrode and buret tip is optimal

(see Installation Instructions for

the Titrando).

"Overshoot"

Electrode responds too

slowly.

Replace the electrode.

Titration time too

long.

Unfavorable control

parameters.

- Select the Titration rate

optimal or fast.

- Select a smaller Dynamics under

User-defined parameters.

- Increase the Max. rate under

User-defined parameters.

- Increase the Min. rate under

User-defined parameters.

Reduce

the

Min. rate under

Control parameters/User-def.

parameters.

Results widely

scattered.

Electrode responds too

slowly.

Replace the electrode.