3.8 Titrants

42

PC Control / Touch Control

mean value of the determinations has been saved as the titer (see

Section 4.5.1).

You can monitor the time interval after which the titrants titer must be

redetermined.

Open the dialog Titer options/Validity and activate the

checkbox for Monitoring.

Enter either the time interval for the validity of the titer or the

date for the next titer determination. When either the time

interval or the date is edited the other value will be updated

automatically.

Select an Action to be carried out when the time interval has

expired. If the titrant is used in the current method then the time

interval will be checked at each START of a determination.

When the monitoring interval has expired one of the following actions

can be selected and carried out:
Display message: a message is shown. You can choose whether

you want to continue with the determination or cancel the run. If you

carry out the run then the fact that the monitoring interval had

expired when the determination was carried out will be documented

in the determination data. The message can be viewed under

Results/More determination data/Messages (see

Section 3.18.2).