4.10.3 Preparing (PREP) and emptying (EMPTY)

Preparation

is used to rinse the cylinder and the tubing. Empty is used

to empty the cylinder and the tubing. The empty command is only

available for Dosino-type dosing devices.
Control device: Selects the control device (Titrando, USB Sample

Processor etc.) to which the dosing device is connected and which

is to be prepared or emptied. This button is only shown when sev-

eral control devices have been configured under System/Device

manager

(see Section 3.10). The device names are always shown in

the selection list.

Dosing device: Dosing device to be prepared or emptied. All dos-

ing device connections (MSB) are always shown.

Titrant: Titrant with which the preparation or empty command is to

be carried out. The titrant can be selected from the list of titrants

which have been defined under System/Titrants. If intelligent ex-

change or dosing units are used then a check will be made in the

method sequence that the correct titrant is present in the connected

dosing device and whether the type of dosing device is correct.

With non-intelligent exchange or dosing units the cylinder volume is

checked. At the start of the command the validity of the titer and the

expiry date of the selected titrant will be checked as well as the GLP

test interval for the exchange or dosing unit. The tubing parameters

defined under System/Titrants for the selected titrant will be used

for preparing and emptying the exchange or dosing unit. The pa-

rameters for the preparation are also defined under Sys-

tem/Titrants. If no titrant has been defined then the standard tub-

ing parameters and the default parameters for the preparation will

be used (see Section 3.8.6).

Cylinder volume to (PREP only): Select the port of the Dosino

where the titrant is ejected. This parameter is only active when the

titrant is not defined. If a titrant has been defined, the cylinder vol-

ume is ejected via the port defined as dosing port for prepar-

ing/emptying (see Section 3.8.6).

Air inlet (EMPTY only): Select the port where the air is aspirated.

This parameter is only active when the titrant is not defined. If a ti-

trant has been defined, the port defined as special port is used (see

Section 3.8.6

).

A detailed description of the preparation and emptying sequences is

given in Section 3.24.2, p. 182 and p. 183).