5 Troubleshooting – Maintenance

PC Control / Touch Control

291

Problem

Possible cause

Remedy

Does your sample evolve

water extremely sluggishly?

Adjust the method, add solubilizer,

work at increased temperature

(poss. use KF oven), see technical

literature.

Are acids esterified in your

sample?

Change the solution more

frequently and increase the buffer

capacity of the solvent.

Does your sample contain

ketones or aldehydes?

Use special reagents suitable for

ketones and aldehydes.

The drift increases

with every titration.

The pH value is no more in

the optimal range.

Add buffer solution, see technical

literature.

The endpoint is

reached "too

rapidly".

Reduce

the

Max. rate under

Control parameters/User-def.

parameters.

The buffer capacity of the

solvent can be depleted with

two-component reagents.

Change the solution.

The titration times

become longer and

longer.

If the drift increases at the same

time, see there.

5.1.6 STAT

titrations

Problem

Possible cause

Remedy

The titration is not controlled

properly.
Rule of thumb for the

minimum rate: expected rate

of reaction in μL/min / 10

-

Reduce the Min. rate under

Control parameters/User-def.

Parameters, possibly the Max.

rate as well.

-

Increase the Dynamics under

Control parameters/User-def.

parameters.

Insufficient stirring.

Increase the stirring rate.

Improper assembly of

electrode and buret tip.

Make sure that the assembly of

electrode and buret tip is correct

(see Installation Instructions for the

Titrando).

The control point is

not kept constant

(The controller

oscillates).

Tubes are kinked.

Use new tubes.