4

Parameters

PC Control / Touch Control

225

Cylinder volume

Max. dosing rate

Max. flow rate

Exchange unit

Dosing unit

Exchange unit

Dosing unit

1 mL

2.25 mL/min

---

approx. 130 mL/h

---

2 mL

---

5.00 mL/min

---

approx. 300 mL/h

5 mL

11.25 mL/min 12.50 mL/min approx. 670 mL/h approx. 750 mL/h

10 mL

22.50 mL/min 25.00 mL/min approx. 1.3 L/h

approx. 1.5 L/h

20 mL

45.00 mL/min 50.00 mL/min approx. 2.7 L/h

approx. 3.0 L/h

50 mL

112.50 mL/min 124.50 mL/min approx. 6.7 L/h

approx. 7.5 L/h

The dosing rate corresponds to the flow rate for continuous dosing.
The parameters for the stirrer are edited under Stirrer:
Stirrer (d): Selects the stirrer to be used in the titration (or meas-

urement, calibration, dosing). All the stirrer connections (MSB) will

always be shown.

Stirring rate (d): You can check the optimal stirring rate manu-

ally (see Section 3.24.3). The stirring direction changes as the sign

changes. The default setting 8 corresponds to 1000 rpm (see Fig.

4.14

).

Switch off automatically (d): This function is used to switch off

the stirrer automatically at the end of the titration (or measurement,

calibration, dosing).

Speed/min

Stirring rate

2000

1500

1000

500

0

-15

-10

-5

-500

-1000

-1500

-2000

0

5

10

15

Fig. 4.14: Setting the stirring rate and speed