2 functions of the individual remote lines – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 320

Advertising
background image

6.2 Remote box

308

PC Control / Touch Control

6.2.2 Functions of the individual remote lines

Name

Pin.

Function

Input

0

21

START

On activation the current method is started.

t

pulse

> 100 ms

Input

1

9

STOP

On activation the current determination is halted

(Stop).

t

pulse

> 100 ms

Input

3

10

CANCEL

On activation the current command in the

determination run will be halted.

t

pulse

> 100 ms

Input

4

23

-

Not occupied, can be used for scanning the

Sample Processor (sample in work position).

Input

5

11

- Not

occupied.

Output

0

5

READY

The Titrando is ready to receive a start signal.

Output

1

18

Cond. OK

Line will be set when conditioning is OK for SET

and KFT titrations. The line remains set until the

titration itself is started with [Start].

Output

2

4

TITRATION

The Titrando is carrying out a data producing

sequence.

Output

3

17

EOD

End of Determination

Pulse (t

pulse

= 200 ms) after a determination or

after a buffer/standard during calibration using

a Sample Processor.

Output

4

3

- Not

occupied.

Output

5

16

ERROR

Line is set for error message display.

Output

6

1

- Not

occupied.

Output

7

2

WARNING

Line is set for warning message display.

Voltage:

+5

V

15

0

V

14/25

I ≤ 40 mA
0 V: active (low), 5 V: inactive

Advertising