Installation

PC Control / Touch Control

5

2 Installation

Connecting the Touch Control to the Titrando is described in the

manual for the Titrando as well as the setup of the titration system with

peripheral devices such as stirrers and dosing devices.

2.1 Setting up the Touch Control

2.1.1 Packaging

The Touch Control and its specially packed accessories are supplied in

very protective special packaging. Please store this packaging in a safe

place; it is the only way in which the safe transport of the instrument can

be guaranteed.

2.1.2 Checks

Immediately after receipt, check whether the shipment has arrived

complete and without damage by comparing it with the delivery note.

2.1.3 Location

The Touch Control has been designed for internal laboratory use and

should not be used in explosion-endangered locations.
Place the instrument on a suitable vibration-free laboratory bench,

protected as much as possible from corrosive atmospheres and

contact with chemicals.
Choose a location where the temperature is usually between +5 °C and

+40 °C. The instrument should be protected against excessive

variations in temperature and direct sunlight.