The parameters for the titration sequence are edited under Titration

parameters

:

Titration direction (d): The titration direction is normally deter-

mined automatically from the initial measured value and the set

endpoint. It is recommended that, whenever possible, a positive or

negative alteration of the measured value is entered. If two end-

points have been set then the titration direction will be defined

automatically. In this case the setting will be ignored.

Extraction time (d): The titration takes place during this time. The

titration will not be stopped until the extraction time has elapsed

(even if the EP has already been reached). The entry of an extrac-

tion time may be advisable, e. g. for the titration of sparingly soluble

samples.

Temperature (d): Manually entered titration temperature. If a tem-

perature sensor is connected and temperature measurement is set

to automatic or continuous under Sensor (see Section 4.1.5) then

the temperature will be measured continuously. The value is used

for temperature compensation in pH measurements.

Time interval MP (d): Time interval for entering a measuring point

in the list of measuring points.

The conditions for terminating the titration are defined under Stop con-

ditions

if this does not occur automatically because either the set

endpoint has not been reached or the stop criterion has not been ful-

filled:
Stop volume (d): Stops when the preset volume has been added

since the start of the titration. You should adapt the stop volume to

suit the size of your titration vessel in order to prevent it from over-

flowing.

Stop time (d): Stops when the preset time has elapsed after the

start conditions have been fulfilled.

If several stop conditions have been set then the first criterion to be ful-

filled will stop the titration.
Filling rate (d): Rate at which the dosing cylinder is refilled after

the titration. The maximum filling rate depends on the cylinder vol-

ume of the exchange or dosing unit used (see Section 3.8.6).

Under Conditioning (d) the conditions for pre- and reconditioning are

defined. With conditioning switched on, the first time that the method is

started only the command to start conditioning will be executed. None

of the other commands will be processed yet. The solution in the titra-

tion cell is titrated to the endpoint with the given control parameters and

is then kept there continuously. The method sequence itself only starts

when [Start] is pressed again.
Start drift (d): As soon as this volume drift is undercut "Condi-

tioning OK" appears and the titration method can be started.

Drift correction (d): The EP-volume can be drift-corrected. With

auto

the volume drift value at the start of the titration will be taken

over automatically, multiplied by the drift correction time and sub-

tracted from the EP-volume. The drift correction time is the time in-