5 output lines – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 113

Advertising
background image

3

Operation

PC Control / Touch Control

101

To input the binary pattern please consult the online help.

3.14.5 Output lines

Binary patterns can be defined for setting remote lines with the CTRL

command (see Section 4.11.2) during the determination sequence or in

manual operation (see Section 3.24.5); these can then be selected

when editing the parameters.

Open the dialog Templates/Output lines.

You can define new templates with [New]. Existing templates can be

removed from the list with [Delete] or edited with [Edit].

List of predefined output lines
Parameter Signal

pattern

Function

Continue *********p****

sends a pulse to continue to the connected Sample

Processor

INIT

00000000000000

initializes the Remote interface

METER Cal C

*********1001*

switches ion meter to Conc calibration

METER Cal pH

*********0101*

switches ion meter or pH meter to pH calibration and

starts it

METER enter

*********1111*

simulates the <ENTER> key for ion meter or pH meter

(essential for pH calibration with 691/780/481, in order to

start the measurement of the second buffer)