5.1 Troubleshooting

294

PC Control / Touch Control

5.1.11 File manager

Problem

Possible cause

Remedy

Files, e. g. methods,

which have been

saved on a data card

by a PC, cannot be

read by Touch

Control.

Windows Explorer has been

used for copying and the

correct folder structure has

not been maintained (see

Section 3.11).

Use PC Control to copy the files.

5.1.12 Miscellaneous

Problem

Possible cause

Remedy

Touch Control

cannot be switched

on.

Touch Control is connected

to an MSB socket of the

Titrando.

Connect the Touch Control to the

Controller socket (see Installation

Instructions for the Titrando).

The PC is connected to an

MSB socket of the Titrando.

Connect the PC to the Controller

socket (see Installation Instructions

for the Titrando).

PC Control does not

recognize the

Titrando.

The suitable USB driver has

not been found.

Remove the USB plug from the PC

and reinsert it. Please note that the

PC Control software must be

installed before the Titrando is

connected to the PC (see

Installation Instructions for the

Titrando).