Dosing unit – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 59

Advertising
background image

3

Operation

PC Control / Touch Control

47

Enter the Volume of reagent which is to be dosed in during a rinsing

cycle. The default value is the cylinder volume, i.e. the contents of

one cylinder is dosed in during a rinsing cycle. You can also enter a

different volume.

Select the number of Cycles to be carried out during the

preparation function. At least two rinsing cycles should be carried

out in order to remove all the air bubbles.

Enter the Dosing rate and the Filling rate to be used during the

PREP function for aspirating and ejecting the reagent. Lower rates

should be entered for substances with higher viscosities.

The maximum dosing rate and maximum filling rate for the

exchange unit depend on the cylinder volume:

Cylinder volume

Max. rate

1 mL

3.00 mL/min

5 mL

15.00 mL/min

10 mL

30.00 mL/min

20 mL

60.00 mL/min

50 mL

150.00 mL/min

You can always enter a value between 0.01 and 166.00 mL/min

independent of the cylinder volume. When the function is carried out the

rate will be reduced automatically to the highest possible value.

Dosing unit
The parameters apply to the preparation (PREP command) and

emptying (EMPTY command) functions. The preparation and emptying

functions can be found under manual control, which can be started

with the [Manual] fixed key (see Section 3.24.2). In the Device

manager, you can define for each dosing device when a message is to

be shown to remind you that the PREP function must be carried out for

the particular dosing device (see Section 3.10).