Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 136

Advertising
background image

3.16 Editing parameters

124

PC Control / Touch Control

Under Memory select the memory location in which the

determinations are to be stored. Determinations can only be stored

on Card 1 or Card 2 or a shared file system (requires 847 USB Lab

Link, see Section 3.10.11). Enter the Group in which the

determinations are to be stored.

If you have retained the standard settings for the File name then it will

be made up of the first 16 characters which you have entered for

Identification 1 and the determination time (date and time):

Identification 1-YYYYMMDD-hhmmss. For the first 16 characters of the

file name you can also select Identification 2 or the name of the

method or you can enter a different text. The date and time will be

appended so that the file name is always unambiguous.
If the Write protection checkbox is activated then the saved

determination files will be automatically write-protected and therefore

protected against accidental alterations or alterations by unauthorized

persons. This function is switched on by default so that the original

data is also retained.
If the Send PC/LIMS report checkbox is activated then an ASCII report

containing all the important data of the determination will be sent

automatically at the end of the determination. The settings for sending

and saving the PC/LIMS report are made in the device manager under

Edit device/PC/LIMS report (see Section 3.10.12). The file name

corresponds to the name of the determination preceded with

PC_LIMS_Report-. I.e. the format is PC_LIMS_Report-Identification 1-

YYYYMMDD-hhmmss.txt if you keep the default settings for the File

name under Save determination automatically. The date and time

correspond to the determination time. A detailed description of the

contents of the PC/LIMS report can be found in the Titrando PC/LIMS

Report Guide.

Advertising