Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 275

Advertising
background image

4

Parameters

PC Control / Touch Control

263

4.8 Calibrating conductivity measuring cells (CAL

Cond)

The cell constant of a conductivity measuring cell can be determined

with the CAL Cond command.
Under Calibration parameters you can edit the parameters for the

calibration sequence.
Signal drift (d): Drift, i.e. the alteration in the measured value per

minute that has to be reached for measured value acceptance. off

means that the measured value acceptance takes place when the

maximum waiting time has elapsed. This may be advisable if the

measuring cell has a very slow response.

Waiting time min. (d): The measured value is not accepted until

the minimum waiting time has elapsed, even if the measured value

drift has already been reached. The minimum waiting time is only

important for drift-controlled measurements.

Waiting time max. (d): If the signal drift is switched off or has not

yet been reached then the measured value will be accepted after

the maximum waiting time has elapsed.

Measuring temp. (d): Manually entered measuring temperature. If a

temperature sensor is connected and temperature measurement is

set to automatic or continuous under Sensor (see Section 4.1.5)

then the temperature will be measured. If off is defined then the

temperature is requested after the start.

Under Standard the characteristics of the calibration standard are en-

tered. These values are proposed in the query after the start of the

method.
Request at calibration start: If this parameter is switched on

then you are requested – after the start of the method – to enter the

following parameters.

Conductivity std. (d): Enter the conductivity of the calibration

standard used.

Reference temp. std. (d): Enter the reference temperature at

which the calibration standard has the conductivity entered above.

Temp. coefficient (d): Enter the temperature coefficient corre-

sponding to the above reference temperature.

Under Stirrer control you can edit the parameters for controlling the

stirrer during the calibration. Take care that you have selected a stirrer

under Edit command/Stirrer.
Stir solution during measurement (d): Automatically switches

the stirrer on/off at the start and the end of the measurement.

Stir before meas. (d): If stirring is switched off during the meas-

urement, the solution can be stirred before the measurement. Enter

the corresponding time here.

Pause before meas. (d): If stirring is carried out before the meas-

urement, a pause interval can be entered here. No stirring or meas-

uring occurs during this time.