3.24 Manual control
190
PC Control / Touch Control
Select the required lift position and press [Assign].
Move to rack position
[Rack pos.] accesses the selection of the special beaker
positions.
Select the required entry under Rack position or enter an absolute
rack position and press [Start]. The required rack position is
moved to.
Only those special beakers will be offered for selection that are defined
in the current rack table (see Section 3.10.5).
The shift rate and the shift direction of the sample rack can be
changed here. If the option "auto" is selected as shift direction, then the
sample rack rotates automatically in the direction in order to result in
the shorter way. "+" means that the direction of rotation is always
counterclockwise and "–" means that the direction is always clockwise.
The swing rate is the speed of the robotic arm.
External positions
With [Ext. pos.] you access the selection of the 4 external
positions.