Exchange unit – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 58

Advertising
background image

3.8 Titrants

46

PC Control / Touch Control

Fig. 3.8: Exchange unit tubing connections

Under Exchange unit/PREP parameters and Dosing unit/PREP

parameters you can enter the parameters for carrying out the

preparation (PREP command) and emptying (EMPTY command;

Dosing units only) functions. The preparation function is used to fill

bubble-free and to rinse the exchange or dosing units if they have not

been used for a long time. When this function has been carried out all

the tubing and the cylinder will be filled with reagent. These parameters

are different for exchange units and dosing units.
You should carry out the preparation function before the first

determination or once per day.

Exchange unit
The preparation function can be found under manual control; which

can be started with the [Manual] fixed key (see Section 3.24.2). In the

Device manager, you can define for each dosing device when a

message is to be shown to remind you that the PREP function must be

carried out for the particular dosing device (see Section 3.10).