3.15 Loading methods

110

PC Control / Touch Control

The following method templates can be selected:

Method template

Command sequence

Application

Dynamic

Titration pH

DET pH (dynamic titration)

CALC

REPORT (result report and

titration curve)

Dynamic

Titration U

DET U (dynamic titration)

CALC

REPORT (result report and

titration curve)

Dynamic reagent addition, for

rapid equivalence point titrations,

large volume increments in the flat

part of the curve; small increments

near the EP, many measuring points

near the EP.

For all standard titrations

(acid/base titrations, precipitation

titrations, redox titrations).

Monotonic

Titration pH

MET pH (monotonic titration)

CALC

REPORT (result report and

titration curve)

Monotonic

Titration U

MET U (monotonic titration)

CALC

REPORT (result report and

titration curve)

Reagent addition with constant

volume increments, relatively few

measuring points near the EP.

For titrations with relatively high

signal variations or a suddenly

occurring potential jump, very flat

potential jumps, slow titrations

(non-aqueous titrations, certain

redox titrations, surfactant

determinations).

Endpoint

Titration pH

SET pH (endpoint titration)

CALC

REPORT (result report and

titration curve)

Endpoint

Titration U

SET U (endpoint titration)

CALC

REPORT (result report and

titration curve)

Titration to a predefined endpoint.

For rapid routine determinations,

titration to particular standards and

titrations in which an excess of titrant

must be avoided (p+m values, pH

adjustment).

Karl

Fischer

Titration

WAIT

REQUEST (data scanning)

KFT Ipol (Karl Fischer

titration)

CALC

REPORT (result report and

titration curve)

Titration for water determination by

Karl Fischer with pre-and post-

conditioning.

STAT pH

without pre-

titration

STAT pH (Titration with

holding the measured value

constant)

EVAL RATE (Rate evaluation)

CALC

REPORT (result report and

titration curve)

Titration to a predefined control

point. The control point is kept

constant by continuous reagent

addition, e. g. for determination of

enzyme activity.

Advertising